FDI-01 Probador de impacto de caída libre

The beginning of the test, choose test method, and estimate an initial mass and Δm.

Start the test.

If the first specimen fails, decrease the mass of the falling dart by Δm.

If the first specimen is not a failure, increase the mass of the falling dart by Δm.

Continue the test according to this rule. In brief, increase or decrease by Δm according to whether the former specimen is a failure or not.

After 20 specimens, calculate the total number of failed specimens N.

If N equals to 10, the test is over.

If N is less than 10, add specimens and continue to test until N equals to 10.

If N is greater than 10, add specimen and continue the test until the number of non-failure specimens reaches 10.

 

Then the tester calculates the test results automatically according to specific formulas.

  • Windows menu interface and touch screen are convenient for the user to set test parameters quickly.
  • 2 test methods of A & B and automatic determination of testing status.
  • Electromagnetic suspension and automatic release of the falling dart reduce the error caused by manual operation.
  • Pneumatic clamping, 2 starting modes of manual and pedal switch and built-in observation light are convenient to the user's operation.
  • Professional software supports multi-unit results display, graphic display of testing process and data export and printing.
  • Equipped with RS232 port and micro printer port which is convenient to the data transmission and PC connection.
  • Support Lystem™ Lab Data Sharing System for uniform and systematic data management.